What should be paid attention to when using the gas mass flowmeter
1. When the working conditions of the gas mass flowmeter change (such as changing the medium, the ambient temperature changes greatly, etc.), the zero position of the meter should be readjusted. We adjust through the zero adjustment hole on the side of the air inlet on the outer cover, and the outer cover can also be opened for adjustment. At the same time, the conduit of the instrument needs to be installed horizontally and calibrated with a level. Otherwise, the influence of the change of working conditions on the zero drift will be increased. The frame must not have vibration or swing, so it is not suitable for use on ships.
2. The linearity of the gas mass flowmeter is related to the size of the range. The larger the flow rate, the worse the linearity, that is, the more serious the nonlinearity. Therefore, the range is generally limited within 0 to 4 standard liters/hour (air) to ensure good performance. linearity. In order to measure a large flow rate while ensuring linearity, the shunt principle can be used to expand the measuring range of the meter. If the bypass tube, Venturi tube, orifice plate are used together, the measuring range can be expanded to tens, hundreds, thousands of standard liters per hour, up to tens of thousands of standard cubic meters.
3. Although the specific heat of the real gas will change with the pressure, and even some gases have relatively large fluctuations, the measurement accuracy of the instrument can still be maintained within a certain range.
4. In the selection of conduit material, in addition to considering corrosion resistance, it is better to choose materials with better thermal conductivity. Taking the measurement of nitrogen as an example, it is also tested in the range of 0~100kgf/cm2 pressure and 0~7 standard liters/hour flow rate. The measurement accuracy of nickel tube is 2~2.5%, while that of stainless steel is 3~4%. (The thermal conductivity of nickel is about 3 times that of stainless steel).
5. Since this type of gas mass flowmeter can only work normally when the specific heat of the gas is relatively stable, when the gas composition is unstable, there is fog in the gas, and the working conditions are close to the liquefaction critical area of the gas, etc., due to The specific heat value is very unstable, so it is not suitable to use this kind of instrument. For example, the critical point of liquefaction of ethylene is 50kgf/cm2 and 9.9℃. When the pressure exceeds 30 kgf/cm2 during the test, the reading of the gas mass flowmeter will start to lose stability.
6. If the gas medium to be measured is changed, it is recommended to re-calibrate. In the instruction manual of the instrument, it is often introduced that the uncalibrated gas flow rate is converted only according to the specific heat of the two gases without recalibration. Although this is simple and convenient, it will cause large errors, especially when working under high pressure. We found that the sensitivity of the meter is not completely proportional to the specific heat, and it should be re-calibrated.
7. The gas mass flowmeter needs to be turned on and warmed up before use. Before the gas mass flowmeter is fully warmed up, the operation of the gas mass flowmeter is unstable. For better models, the boot warm-up time is within two hours.
8. During use, when the gas flow rate changes suddenly, the temperature in the tube must be redistributed through heat transfer, so it takes a certain amount of time for the output signal to re-stabilize. In order to reduce this hysteresis, the manufacturer often adds a differential network to the electrical circuit of the instrument to make the output signal respond quickly. This is especially necessary when cooperating with other instruments for automatic flow control.
mass flow meter tuning fork liquid density meter, as the name suggests, find extensive use in low flow turbine flow meter institutions. Since rosemount coriolis mass flow meter has become much dependent on technology in today's world, there is wide use of such mass flow meter.
The guiding vision of Beijing Sincerity Automatic Equipment Co., Ltd is 'Bringing the best to everyone we touch'. By 'The best', we mean the best products, the best people and the best ideas.
The team of engineers and developers at Beijing Sincerity Automatic Equipment Co., Ltd are the best in their own way and we promise to provide timely service to our esteemed clients.
Quick Links
Contact Us
● Add:No. 8 building, 3rd. district,I-Town, Gaoli Zhang
road,Haidian District, Beijing, China 100095
● Contact :Zoe Pan
Skype: testifypp
● Mobile:+86 13601110505
● Whatsapp : +86 18600270515
● Email:info@bjsincerity.com